Linac Photon Beam Calibration Using Virtual Simulator Program

Authors

  • Niko Hysa
  • Erjon Spahiu University of Tirana, Faculty of Natural Sciences, Department of Physics, Tirana, Albania.
  • Dafina Xhako
  • Partizan Malkaj
  • Theodor Karaja

Abstract

In radiotherapy, medical physicists give a major contribution to the safe and effective radiation treatment for patients with cancer. Megavoltage linac photon outputs are determined using the IAEA TRS-398 code of practice or AAPM TG-51 and the results are compared. Beam calibration means determination of absorbed dose to water per 100 monitor units in a water phantom at reference conditions. The measured dose Dw, Q in water at the reference point is a primary parameter for planning the treatment monitor units (MU). Traceability of dose accuracy therefore still depends mainly on the calibration factor of the ion chamber/dosimeter provided by the accredited laboratories. Our data, therefore, imply that the dosimetry level maintained fo the clinical use of linear accelerator photon beams are within recommended levels of accuracy, and uncertainties are within reported values. However, in Albania, the frequent problem is related to resources with respect to both, qualified teachers and equipment, that are at disposal for teaching and training. The concepts of e-learning methods using different non-commercial software, contribute to overcoming this problem. In our case, we use an academic education method to practice radiation oncologists and medical physicists for LINAC beam calibration using a virtual simulator program and Matlab. Our group, after some experiences in calculation methods using Matlab, is focused on a PC based program which simulates the required equipment, the measurement set-up, and the measurement itself. All procedures are modeled according to the IAEA Code of Practice, TRS 398.

Keywords: Radiotherapy, linac, calibration, beam, protocol, virtual simulator program.

Downloads

Published

2016-12-20

How to Cite

Hysa, N., Spahiu, E., Xhako, D., Malkaj, P., & Karaja, T. (2016). Linac Photon Beam Calibration Using Virtual Simulator Program. ANGLISTICUM. Journal of the Association-Institute for English Language and American Studies, 2(4), 353–360. Retrieved from https://www.anglisticum.org.mk/index.php/IJLLIS/article/view/1310

Issue

Section

Volume 2, No.4, August, 2013